t r o u b l e s h o o t i n g
g
Random or single Hct Sensor error messages
•
If the error occurred during calibration, re-calibrate the
analyzer.
•
If the error occurred during sample analysis, run the sam-
ple again.
•
If the error remains, perform
DAILY CLEANER
twice and repeat calibration/analysis.
•
If the error remains, remove the sensor module from the
analyzer and examine the exterior surface of the sensor
module and Hct sensor for liquid or salt deposits.
•
Examine the sensor module’s electronic connector for cor-
rosion.
•
Inspect inside the sensor module for liquid and/or salt
deposits.
•
Make sure all sensor module and electrode surfaces are
dry.
•
Re-install the sensor module, perform
DAILY
CLEANER
and repeat calibration.
•
If failure continues, replace the sensor module.