When the installed reagent module has less than 10% remaining, the
HOME MENU
display will flash the % remaining. Prepare to install
a new reagent module when the % remaining reaches zero.
When all components are installed, confirm that the access door
is closed.
WARMING UP
appears at the bottom of the
HOME
MENU
display. When the sensor module reaches 37.0°C and the
valve module reaches 39.0°C (approximately 30 minutes from a cold
start) the
WARMING UP
message disappears. Calibration can now
be performed. Refer to Calibrate in Operating the Analyzer for further
instructions.
New electrodes require a short stabilization period to prevent drift
errors. This stabilization is minimized by the analyzer, which performs
calibrations every hour for 8 hours. After this stabilization period,
the analyzer will calibrate at the calibration frequency selected under
CAL FREQUENCY
in the
CONFIGURATION
section of the
SETUP MENU
.