t r o u b l e s h o o t i n g
Electrode error messages for more than one electrode
may indicate a flow problem.
•
Remove the electrodes from the sensor module.
•
Inspect the O-ring connections, and replace any cracked O-rings.
•
Inspect the sensor module surfaces that surround the electrodes for
salt deposits and/or liquid.
•
Inspect above and below the compression plate, and the area
around the fluid path exit port.