Masimo
Trends
From any of the capnography parameter setting screens, touch
Trends
, and then change any of the following options:
Options
Description
Factory Default Settings
User Configurable Settings
Y-Axis Max
EtCO
2
,
FiCO
2
The Trend Max is the highest
value displayed.
8.0 vol%
8.0 kPa
60 mmHg
0.1 to 25.0 vol% in steps of 0.1
0.1 to 30.0 kPa in steps of 0.1
1 to 225 mmHg in steps of 1
RR
35
1 to 150 in steps of 1
Y-Axis Min
EtCO
2
,
FiCO
2
The Trend Max is the highest
value displayed.
0.0 vol%
0.0 kPa
0 mmHg
0.0 to 24.9 vol% in steps of 0.1
0.0 to 29.9 kPa in steps of 0.1
0 to 224 mmHg in steps of 1
RR
0
0 to 149 in steps of 1