End-tidal Carbon Dioxide (EtCO2)
From the
End-Tidal Carbon Dioxide Settings
screen, access the following screens:
EtCO2 Alarms
on page 75
About Parameter Information
on page 80
Trends
on page 81
EtCO2 Alarms
From the
End-Tidal Carbon Dioxide
screen, touch
Alarms
, and then change any of the following options:
WARNING:
Alarm limits on the Radius PCG device and the Masimo device are set independently of each other. If
alarm limits are not set the same, the Masimo device may not alarm at the same time as the Radius PCG device. See
Chapter 4: Operation
on page 27 for information on adjusting alarm limits on the Radius PCG device.
Options* Description
Alarm
Priority
Factory Default
Settings
User Configurable Settings
High
Limit
High Limit is the upper
threshold that triggers an
alarm.
Medium
7.9 vol%
7.9 kPa
60 mmHg
0.2 to 24.9 vol% in steps of 0.1, or Off
0.2 to 29.9 kPa in steps of 0.1, or Off
2 to 224 mmHg in steps of 1, or Off
When set to Off, alarm is disabled
Low
Limit
Low Limit is the lower
threshold that triggers an
alarm.
Medium
1.6 vol%
1.6 kPa
12 mmHg
Off, or 0.1 to 24.8 vol% in steps of 0.1
Off, or 0.1 to 29.8 kPa in steps of 0.1
Off, or 1 to 223 mmHg in steps of 1
When set to Off, alarm is disabled