51
Masimo
During immunity testing the essential performance of Radius PCG was assessed based upon the gas measurement
accuracy (see Citation [4]) including gas reading alarm conditions, or generation of technical alarm conditions. For
higher immunity conditions above 100 V/m at frequencies between 4 GHz to 8 GHz (i.e., exposure to rotational radar),
the essential performance was determined based upon the return of the device to normal operation without
permanent damage.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test
Test
Level
Compliance
Level
Electromagnetic Environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV
contact
±15 kV air
± 8 kV
contact
±15 kV air
Floors should be wood, concrete, or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least
30%.
Power frequency
(50 / 60 Hz)
magnetic field.
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at levels characteristic of
typical location in a typical hospital environment.
Radiated RF
IEC 61000-4-3
10 V/m
(80 MHz
to 2.7
GHz)
10 V/m
Portable and mobile RF communications equipment should be used
no closer to any part of Radius PCG than the recommended
separation distance calculated from the equation applicable to the
frequency of the transmitter. Otherwise, degradation of the
performance of this equipment could result.
d = 0.6*SQRT(P) 80 MHz to 2.7 GHz
where P is the maximum output power rating of the transmitter and
d is the recommended separation distance in meters (m).