Masimo
Data Output
Specifications
Adult/Pediatric
Dead space 6 ml, Flow resistance < 0,3 cm H2O (@ 30 LPM)
Infant
Dead space 1 ml, Flow resistance < 1,3 cm H2O (@ 10 LPM)
Gas Analyzer
Specifications
Warm-up
In operation and full accuracy within 15 seconds.
Warm-up time after storage
at -40°C
A warm-up period of 15 minutes is required before Radius PCG is ready for use after
being stored with batteries mounted at −40°C when the ambient temperature is 20°C.
Cool-down time after storage
at 70°C
In operation and full accuracy within 15 seconds.
Calibration
No routine calibration is required.
Total System Response Time < 0.7 seconds