Note:
To preserve and maximize battery life, make sure alarm conditions are addressed promptly and the unit is
turned off when not in use, especially prior to storage.
Note:
Bluetooth functionality is implemented differently by the various connected-device manufacturers and
incompatibility between the Radius PCG and your external device may occur.
Note:
A trained medical professional must determine the proper EMMA Airway Adapter model for each patient
application. No hardware or software configuration changes result from the EMMA Airway Adapter model selected.
Note:
The alarm limits will be reset to default values after Radius PCG powers off.
Cleaning and Service Warnings and Cautions
WARNING:
Properly use and dispose of batteries or they may leak or explode.
WARNING:
Lithium batteries may present a fire or chemical burn hazard if mistreated. Do not disassemble, heat
above 100°C (212°F) or incinerate. Dispose of used cells promptly. Keep away from children.
WARNING:
Use only Alkaline batteries or Energizer Ultimate Lithium L92 batteries. Use of other Lithium batteries
may present a risk of fire or explosion.
WARNING:
Replace batteries immediately when the Battery Status Indicator starts blinking. Remaining battery time
depends on battery type and other circumstances and cannot be reliably predicted.
CAUTION:
Remove alkaline batteries when the Radius PCG will not be in use for more than 30 days to avoid damage
to the device due to batteries that may leak.
CAUTION:
Replace both batteries at the same time to avoid mixing fully and partially charged batteries. These
actions may cause the batteries to leak; resulting in possible damage to the device.
CAUTION:
Do not immerse Radius PCG in any liquid.
CAUTION:
Do not apply excessive pressure on the IR-windows.
CAUTION:
Never saturate Radius PCG completely with any disinfection solution.
CAUTION:
Only perform maintenance procedures specifically described in the manual; otherwise, return Radius PCG
for servicing. Improper maintenance may result in damage to the internal parts. Damage to internal parts may result
in no or inaccurate readings.