![Masimo Radius PCG Operator'S Manual Download Page 89](http://html1.mh-extra.com/html/masimo/radius-pcg/radius-pcg_operators-manual_1735249089.webp)
www.masimo.com
87
Masimo
Alarm/Exception
Messages
Activation
Alarm
Priority
Next Steps
Capnography
Disconnected
•
Radius PCG is out of Bluetooth range.
•
Radius PCG has turned off.
High
•
Ensure Radius PCG is within
Bluetooth range during
operation.
•
Turn Radius PCG on.
Unable to handle
Capnography
channel on
Bluetooth
Radius PCG is paired to Root and a
capnography device is connected to Root
through the MOC-9 connection. The
Capnography
window is populated from the
MOC-9 capnography device.*
High
Only one (1) capnography device can
be connected to Root at one time.
Disconnect one of the capnography
devices.
Unable to handle
Capnography
channel on MOC-9
port X (1
-3)
A capnography device is connected to Root
through the MOC-9 connection and Radius
PCG is paired to Root. The
Capnography
window is populated from the Radius PCG.
High
Only one (1) capnography device can
be connected to Root at one time.
Disconnect one of the capnography
devices.
* Radius PCG does not reconnect automatically after disconnecting the capnography device from the MOC-9
connection. Radius PCG must be disconnected and paired again to communicate with Root and display its data on the
Main Screen
.
Troubleshooting Radius PCG with Root
Symptom
Possible Cause
Correction
Root Bluetooth settings
do not display the ability
to pair Radius PCG.
Root software requires updating.
Upgrade software to V2.0.9.6 or higher.