If no buttons have been activated for a short period during alarm limit adjustment, the Radius PCG will automatically
resume normal operation.
Low EtCO2 Alarm
1.
Press and hold the Power On button until the display shows the “Lo ETCO2
Screen” and the EtCO
2
display shows the current low EtCO
2
alarm limit. See
Controls
on page 31.
2.
Release the Power On button.
3.
To adjust the alarm limit: press the Alarm Silence button (
▲
) to increase,
or the Power On button (
▼
) to decrease the value. It is possible to switch
off the low EtCO
2
alarm by adjusting the limit down to 0. The Radius PCG
will indicate this setting by showing "- -" on the EtCO
2
display during the
adjustment routine.
If no button has been activated for a short period, the Radius PCG will
automatically resume normal operation.