88
Masimo
Symptom
Possible Cause
Correction
Cannot pair Radius PCG to
Masimo device.
•
Masimo device Bluetooth
may be off.
•
Radius PCG may be out of
Bluetooth range with the
Masimo device.
•
Internal failure.
•
Ensure Bluetooth on the Masimo device is
turned on.
•
Ensure Radius PCG is within range of the
Masimo device when pairing.
•
Radius PCG requires service. See
Service and
Return Procedure
on page 63.
Root Specifications
Display Range
Measurement
Display Range*
EtCO
2
(End-Tidal CO
2
)
0% to 25%
0 kPa to 32.5 kPa
0 mmHg to 244 mmHg
FiCO
2
(Fractional Concentration of Inspired CO
2
)** 0% to 25%
0 kPa to 32.5 kPa
0 mmHg to 244 mmHg
RR (Respiration Rate)
0 BPM to 150 BPM