Chapter 8: Service and Maintenance
The following chapter contains information about cleaning, battery operation, performance verification, service,
repair, and warranty.
Cleaning
Radius PCG Cleaning
CAUTION:
Do not immerse Radius PCG in any liquid.
CAUTION:
Do not apply excessive pressure on the IR-windows.
Note:
Before cleaning Radius PCG, remove the batteries and make sure the battery cover is re-attached correctly.
To clean Radius PCG, follow the instructions below:
1.
Remove the Airway Adapter.
2.
Wipe each of the outer surfaces twice or until the surfaces are free of any visible residue, using one of the
following solutions:
•
A cloth moistened with 70% Isopropyl Alcohol
•
A quaternary ammonium chloride solution wipe (for example CaviWipes™)
Note:
Pay particular attention to crevices and hard to reach areas of the device. Use a soft bristled brush
to gently remove any visible residue from crevices as necessary.