Masimo
Barometric pressure
[mmHg]
Radius PCG EtCO
2
readings should be between
5% CO
2
[mmHg]
5% CO
2
[kPa]
660-679
31-36
4,1-4,8
680-699
32-37
4,3-4,9
700-719
33-38
4,4-5,1
720-739
34-39
4,5-5,2
740-759
35-40
4,6-5,4
760-779
36-41
4,8-5,5
780-799
37-42
4,9-5,6
If the unit is reading within the above range then your Radius PCG has been successfully verified.
If the unit is not reading within the above range, disconnect the Airway Adapter from the gas cylinder, and perform a
Zeroing procedure and then repeat the Gas span check procedure. See
Zeroing Procedure
on page 60. If verification
still fails, contact your local distributor for further instructions.