Parameter Settings
The following information contains default alarm limits and information for adjusting EtCO
2
alarm limits.
Default Alarm Limits
The default factory settings for the RR and the EtCO
2
alarms are as follows:
Lower Limit
Upper Limit
RR (No Breath)
3 bpm (20 seconds)
NA
EtCO
2
OFF
50 mmHg (7.0 kPa)
EtCO2 Settings
The adjustment ranges for the EtCO
2
alarm limits are as follows:
Lower range
Upper range
EtCO
2
(mmHg)
OFF; 1 – 89 mmHg
11 – 99 mmHg; OFF
EtCO
2
(kPa)
OFF; 0.1 – 8.9 kPa
1.1 – 9.9 kPa; OFF
Note:
The alarm limits will be reset to default values after Radius PCG powers off.
If the high EtCO
2
limit is decreased close to the low EtCO
2
limit, the low limit will be automatically adjusted in order to
maintain a minimum difference of 10 mmHg (1.0 kPa) between the high and low alarm limit. Similarly, if the low
EtCO
2
limit is increased close to the high EtCO
2
limit, the high limit will be automatically adjusted to maintain a
minimum difference of 10 mmHg (1.0 kPa) between the high and low alarm limit.