Masimo
RR Settings
From the
Respiratory Rate Settings
screen, access the following screens:
RR Alarms
on page 79
About Parameter Information
on page 80
Trends
on page 81
RR Alarms
From the
Respiratory Rate
screen, touch
Alarms
, and then change any of the following options:
WARNING:
Alarm limits on the Radius PCG device and the Masimo device are set independently of each other. If
alarm limits are not set the same, the Masimo device may not alarm at the same time as the Radius PCG device. See
Chapter 4: Operation
on page 27 for information on adjusting alarm limits on the Radius PCG device.
Options* Description
Alarm
Priority
Factory Default
Settings
User Configurable Settings
High
Limit
High Limit is the upper
threshold that triggers an
alarm.
High
50
2-149 in steps of 1, or Off
When set to Off, alarm is disabled
Low
Limit
Low Limit is the lower
threshold that triggers an
alarm.
High
6
Off, or 1-148 in steps of 1
When set to Off, alarm is disabled
High
Caution
Range
Area of the display range that
provides a caution indicator.
NA
Off
Off, or 1 to 15 in steps of 1
When set to Off, alarm is disabled