![Manning YB8 Operating And Instruction Manual Download Page 92](http://html1.mh-extra.com/html/manning/yb8/yb8_operating-and-instruction-manual_701729092.webp)
Clean around inside of pump to remove particulate
matter which can decrease pump tube life
MAINTENANCE
MODEL YB8
April 1999/Manning Environmental Inc.
Page C-2
Inspection
Discharge
Every Month
3 min
Examine the discharge tubing for build up of organic and
Maintenance
Tubing
particulate matter. If there is build up, follow the same
procedure for cleaning the intake tubing. Replace as necessary.
Inspection
Electronics
Every Month
2 min
Clean as needed with warm water and very mild soap
Maintenance
Enclosure
Inspection
Keypad
Every Month
2 min
Clean as needed with warm water and very mild soap. Harsh
Maintenance
abrasive cleaning products can damage the keypad and scratch
the clear window to the display
Inspection
Sealing
Every Year
5 min
Unscrew the electronics enclosure from the refrigeration
Maintenance
Gasket
system and ensure that the gasket is free from debris and is
maintaining its shape and consistency. If the gasket is
exhibiting signs of wear, contact the Manning service
department for information on replacement. Under normal
operating conditions, the gasked should provide many years of
life
Inspection
Sample
As often as
15
The sample bottles should be checked frequently to ensure that
Maintenance
Bottles
needed
min
they are clean. EPA cleaning protocol should be used in the
cleaning of the bottles.
Maintenance
Z-Rust
Every Year
8 min
Once a year, the Z-rust capsule which is used to absorb
Capsule
moisture (in the form of humidity) should be replaced
Maintenance
Lithium
Every 3-5 years 12
The sampler contains an internal lithium battery which should
Battery
min
be replaced every 3 to 5 years
Peristaltic Pump
The peristaltic pump used in the Manning Environmental Inc. YB8 is designed for long life and trouble free
service. The following is a list of routine maintenance items:
Summary of Contents for YB8
Page 2: ......
Page 4: ......
Page 22: ...INSTALLATION AND OPERATION MODEL YB8 April 1999 Manning Environmental Inc Page A 12...
Page 30: ......
Page 61: ...MODEL YB8 PROGRAMMING SECTION April 1999 Manning Environmental Inc THIS PAGE BLANK...
Page 90: ......
Page 106: ......
Page 108: ......
Page 119: ...9 YB8 Sampler Theory of Operation Figure 8 YB8 Sampler Block Diagram...