INSTALLATION AND OPERATION
MODEL YB8
April 1999/Manning Environmental Inc.
Page A-16
Whether the sample event is triggered by a flowmeter or by a time interval, the actual sampling cycle is the
same. For the multiple bottle option, the first action is the advance of the spout to the next bottle. (For single
bottle samplers, this step is omitted.) Next, the sampler turns on the peristaltic pump. The pump begins
rotating counter-clockwise causing air to be forced out of the intake tubing. This clears the intake hose of
any contents or obstructions that may inhibit proper sample collection. After the set amount of purge time
has elapsed, the unit will then reverse the pump so that the rollers are now moving in a clockwise direction.
This causes vacuum to be created in the tube, which causes the source liquid, to begin traveling up the intake
line. If the unit was programmed to rinse, as the source liquid rises in the intake line it will reach the liquid
sensor. The sampler will immediately recognize that fluid has reached the inlet to the pump. It will
instantaneously reverse the direction of the pump (rollers will be moving in a counter-clockwise rotation),
sending the water that had been drawn up back out of the intake line. This in effect rinses the line. If the unit
was not set for a rinse the above steps wil be ommited. Upon completion of the last rinse, source liquid will
again be drawn up the intake line. The system monitors the flow of liquid and when the preset amount has
passed through, the pump will reverse operation again (rollers moving counter-clockwise). This purges
excess fluid out of the pump and clears the intake line. Depending on how the unit is programmed or
configured, after completing the post sample purge, the sampler will now stop operation or continue
performing those functions which it has been programmed. For a complete description of programming the
sampler see the programming section in this manual.
If the multiple bottle option is being used, the distribution spout remains stationary until the next sample
event. This delay prevents cross-contamination of the next sample.
Sample Recovery
Immediate sample recovery is not required since the sampler will automatically shut down when the sample
container is full (single bottle only), a pre-set number of samples have been taken, or when the program is
complete. However, sample analysis may require quick recovery to maintain sample freshness or to add
chemicals.
If the intent is to leave the containers in the suspension plate, caps can be installed over the suspension
collars. Remove the suspension plate (with bottles) from the refrigerator. Lift the distribution assembly off
the suspension plate and place it on the wire racks mounted in the refrigerator.
It may be easier to remove the distribution assembly first, and then install the bottle caps. To seal the 350ml
glass bottles, replace cap liners, then place caps on bottles.
Summary of Contents for YB8
Page 2: ......
Page 4: ......
Page 22: ...INSTALLATION AND OPERATION MODEL YB8 April 1999 Manning Environmental Inc Page A 12...
Page 30: ......
Page 61: ...MODEL YB8 PROGRAMMING SECTION April 1999 Manning Environmental Inc THIS PAGE BLANK...
Page 90: ......
Page 106: ......
Page 108: ......
Page 119: ...9 YB8 Sampler Theory of Operation Figure 8 YB8 Sampler Block Diagram...