MODEL YB8
PROGRAMMING SECTION
April 1999/Manning Environmental Inc.
3DJH%
Multiple Samples per Bottle
Multiple Samples per Bottle is not a stand alone function but works in conjunction with other TIME, FLOW
and certain * Modes (*02, *04, *05, *06, *09) to expand the capabilities of the sampler. The sampler places
from 2 to 99 samples in each bottle. In order to use Multiple Samples per Bottle, the user selects the mode of
choice - a TIME, FLOW, or * Mode. After entering the required information, the PUSH START/OPTIONS
prompt will appear on the display. At this cue press <MULTI SAMPLE> and enter the number of samples
per bottle. For example, if Flow Mode were being used and Multiple Samples Per Bottle is set at 5, each time
an event occurred, such as a contact closure or the end of a time interval, the sampler would place a sample in
a bottle. When the next event occurs, the sampler would place another sample in the SAME bottle, until 5
samples had been placed in that bottle. It would then advance the spout to the next bottle in sequence. The
sampler would then repeat the process above for the current bottle. This would continue until the total
number of bottles the unit is configured for (set in *99) have received their allocation of samples.
Display on LCD
Explanation
SAMPLER READY
04/30
04:30:02
This display indicates the sampler is ready to program. Multiple
Samples per Bottle is not a stand alone program. It works in
conjunction with FLOW, TIME and certain * Modes. In this
example <FLOW> was pushed as the mode of choice.
PUSH START/OPTIONS
04:30:02
At this prompt the user selects Multiple Samples per Bottle, by
pressing <MULTI SAMPLE>.
SAMPLES PER BOTTLE?
__ __
04:30:02
Input the number of bottles into which 1 sample will be placed during
a sampling event and press <ENTER>.
PUSH START/OPTIONS
04:30:02
At this point, simply press <START>. NOTE: Multiple Bottles per
Sampling Event and Multiple Samples per Bottle cannot be selected
simultaneously. They are mutually exclusive options.
FLOW MODE
__ __ __ __
04:30:02
The unit is now waiting for a contact closure to initiate the sample
sequence.
Summary of Contents for YB8
Page 2: ......
Page 4: ......
Page 22: ...INSTALLATION AND OPERATION MODEL YB8 April 1999 Manning Environmental Inc Page A 12...
Page 30: ......
Page 61: ...MODEL YB8 PROGRAMMING SECTION April 1999 Manning Environmental Inc THIS PAGE BLANK...
Page 90: ......
Page 106: ......
Page 108: ......
Page 119: ...9 YB8 Sampler Theory of Operation Figure 8 YB8 Sampler Block Diagram...