PROGRAMMING SECTION
MODEL YB8
April 1999/Manning Environmental Inc.
3DJH%
Sampler Configuration Functions
There are 3 major configuration functions the user must be concerned with (*99, *20, & *19). Step-by-step
programming instructions and descriptions of each function are detailed on the pages that follow :
*99 Sampler Set-Up
*99 allows the user to set the sampler’s configuration. For proper operation, it is critical the unit is correctly
configured. The memory comes preset with the sampler’s defaults. These defaults can be reviewed or
changed by entering the configuration mode (explanations and step-by-step instructions are given below).
Once entries have been made in *99, re-entering the configuration mode is not necessary unless changes in the
data are needed.
Display on LCD
Explanation
SAMPLER READY
04/30
04:30:02
This display shows the sampler is ready to program. It displays the
current time. From here the user can enter any TIME, FLOW, or *
Mode. Press the * key to access the * Mode.
ENTER * MODE?
__ __
04:30:02
Prompts the user to enter either a program or the configuration
function. Press 99 and <ENTER> to configure the sampler.
SAMPLER SETTING?
__
04:30:02
Sets the sampler to a specific type of operation or bottle
configuration:
1 = Single Bottle
2 = Multi-Bottle
Other numbers are not valid and will cause the sampler to
malfunction. Enter the desired configuration and press <ENTER>
# OF BOTTLES?
__ __
04:30:02
Sets the number of bottles (1,2,3,4,6,8,12, or 24) in the sampler. If 1
was selected for the SAMPLER SETTING above, this prompt is
bypassed. Input the number of bottles and press <ENTER>.
SAMPLE VOLUME?
__ __ __ __
04:30:02
Sets the sample volume (in milliliters), to be collected per sample.
TUBE LENGTH?
__ __ __ __
04:30:02
Records the length
(1 to 99 feet)
of the sample intake tube.
DRAW HEIGHT?
__ __ __ __
04:30:02
Sets the draw height or lift height
(1 to 25 feet)
of the sampler.
Summary of Contents for YB8
Page 2: ......
Page 4: ......
Page 22: ...INSTALLATION AND OPERATION MODEL YB8 April 1999 Manning Environmental Inc Page A 12...
Page 30: ......
Page 61: ...MODEL YB8 PROGRAMMING SECTION April 1999 Manning Environmental Inc THIS PAGE BLANK...
Page 90: ......
Page 106: ......
Page 108: ......
Page 119: ...9 YB8 Sampler Theory of Operation Figure 8 YB8 Sampler Block Diagram...