INSTALLATION AND OPERATION
MODEL YB8
April 1999/Manning Environmental Inc.
Page A-2
4. Program the Y
B
8.
Hardware
Functional Specifications:
Size
Control Unit: 10.75" (27.305cm)W x 7" (17.78cm)H x 9.75" (24.765cm)D
Refrigeration Unit: 23.875" (60.65cm)W x 34.5" (87.63cm)H x 24" (60.96cm)D
Total Unit: 23.875" (60.65cm)W x 41.5" (105.41cm)H x 24" (60.96cm)D
Weight
Dry Weight: 110 lb (49.89 kg) with refrigerator.
Environmental
Nema 4X, 6 housing around electromechanical components.
Protection
Sample Cooling
Industrial Grade refrigeration unit.
Temperature
0°C to 50°C (32°F to 122°F) without optional enclosure and heater.
Limits
Sample Pump
High Speed peristaltic, dual roller design with impact and corrosion resistant Delrin
Safety
Integral safety kill switch. Ensures when the clear plastic face plate is removed from
®
plastic pump body.
the pump, it prevents accidental powered rotation.
Pump Tubing
3/8" ID by 5/8" OD by 1/8" thick medical grade silicone rubber pump tubing.
Tube Life
Recommended maximum of 1,000,000 pump revolutions based on a standard sample.
A standard sample equates to 5ft of head, 10 foot PVC intake tube, and 200 ml sample
size.
Maximum Lift
28 ft (8.5344 m).
Transport
Minimum of 3 ft/s at 3 ft of lift (0.9144 m/s at 1 m) and 2.0 ft/s at 20 ft of lift (0.6096
Velocity
m/s at 6.1 m).
Sample Volume
Programmed directly in increments of 1 milliliter up to a maximum of 9,999 ml.
Accuracy
± 10ml or ± 10% of the programmed volume, whichever is greater.
Repeatability
± 5ml or ± 5% of the average largest and smallest sample volume in a sample set,
whichever is greater.
Liquid Sensor
Continuity type or Ultrasonic (optional)
Summary of Contents for YB8
Page 2: ......
Page 4: ......
Page 22: ...INSTALLATION AND OPERATION MODEL YB8 April 1999 Manning Environmental Inc Page A 12...
Page 30: ......
Page 61: ...MODEL YB8 PROGRAMMING SECTION April 1999 Manning Environmental Inc THIS PAGE BLANK...
Page 90: ......
Page 106: ......
Page 108: ......
Page 119: ...9 YB8 Sampler Theory of Operation Figure 8 YB8 Sampler Block Diagram...