The serial output from the sampler is fixed at 9600 baud, 8 bits, no parity and 1 stop bit. The
data log is output as ASCII text, comma delimited. *91, option 4 is used to download the data log from
the sampler. If using a computer to record the data log output from the sampler, a terminal program
must be running on the computer.
Sampler Alarm Option.
The alarm option provides three sets of normally open relay contacts that are closed when the
sampler detects that certain conditions exist. The relay contacts are rated 5 Amps at 230VAC or
30VDC. A 3-foot (1 Meter) cable connects to the female panel connector on the sampler chassis to
allow external alarm connections.
The alarm cable is terminated with six 18-gauge wires in a pigtail. To connect the alarm cable to
the sampler, line up the connectors and carefully push the cable connector into the panel connector.
Then, secure the cable by tightening the threaded shell. See Figure 4.
SAMPLER
CHASSIS
FEMALE PANEL
CONNECTOR
MS818013 ALARM CABLE
(3 FT/1 M)
SEE CHART FOR
WIRING DETAILS
Note: Pin 7 (center) is not used
Female Panel Connector Detail
Figure 4.
Pin #
Wire Color
Function
1
Black
Sample Cycle
2
Red
3
White
Missed Sample
4
Green
5
Orange
Bottle Full/ End of
Sequence
6
Blue
Summary of Contents for YB8
Page 2: ......
Page 4: ......
Page 22: ...INSTALLATION AND OPERATION MODEL YB8 April 1999 Manning Environmental Inc Page A 12...
Page 30: ......
Page 61: ...MODEL YB8 PROGRAMMING SECTION April 1999 Manning Environmental Inc THIS PAGE BLANK...
Page 90: ......
Page 106: ......
Page 108: ......
Page 119: ...9 YB8 Sampler Theory of Operation Figure 8 YB8 Sampler Block Diagram...