sample through the standard addition. Enter the confidence interval of the expected value in
the “Confidence interval” field using the alphanumeric keypad.
Enter the LOT for test kit and standard on the alphanumeric keypad and confirm with
to
open the measurement window. The user-defined standard will not be saved in the list of
standards until the measurement has been carried out.
Figure 101: Entering the data of a spike addition measurement
After inserting and measuring the two sample solutions, the result of the measurement will be
displayed as shown in Figure 102. In the upper right corner
the result will be rated “Passed”
(green) or “Failed” (red). The measured value is compared against specified confidence
interval. In addition to the measurement results for the original sample and the spiked
sample, the calculated increase in concentration and the expected value are also displayed.
Figure 102: Result of a spike addition measurement