System
64
NANOCOLOR
®
UV
/
VIS
II and
NANOCOLOR
®
VIS
II V 1.1 / 10.2017
Figure 75: System menu
6.2.1
System info
Press the
icon to open a window showing the instrument data (Figure 76). This is where
information on the manufacturer, instrument model, serial number, operating system version,
software and firmware version, and the number of system starts are listed. Please have this
information ready when contacting the MACHEREY-NAGEL technical support.
Figure 76: System info
6.2.2
System check
Press the
icon to open the dialogue for the system check menu. This is where a periodic
check of the system can be arranged by marking the checkbox. When activated, it will be
performed every 180 minutes following the latest measurement
.
The system check can be triggered manually by pressing
. A lamp test, a wavelength
Summary of Contents for NANOCOLOR UV/VIS II