IQC
71
NANOCOLOR
®
UV
/
VIS
II and
NANOCOLOR
®
VIS
II V 1.1 / 10.2017
When the deuterium lamp is tested, please remember that it has to be heated up
beforehand, which takes about 90 seconds.
6.6.1.2
NANOCHECK
The NANOCHECK program is used to check the photometric accuracy of the
spectrophotometer. For this purpose, you will need the
NANOCONTROL
NANOCHECK test
solutions (REF 925 701) offered by MACHEREY-NAGEL. Press the
icon to open the
NANOCHECK program menu (see Figure 87). In order to enter the wavelength, reference
value, and confidence interval data from the product certificate, each field must be manually
selected and the corresponding value entered via the alphanumeric keypad. The data for the
input NANOCHECK lot will be saved automatically after the program is run for the first time.
Alternatively, a pre-programmed LOT can be chosen from the list box for the NANOCHECK
LOT via the
icon. This list also includes all of the LOTs entered manually. To delete a
LOT with all of its associated data, press and hold down that entry for a few seconds. Press
the
icon that appears to delete the entry from the list.
Figure 87: Entering the desired values for
NANOCONTROL
NANOCHECK test solutions
After selecting the desired LOT or entering all necessary data, the measurement is started by
pressing the
icon. An automatic menu guide will ask you to insert the NANOCHECK test
solutions (see Figure 88). The progress of the test is shown by the advancing progress bar.