accuracy test, a filter test, and detector test will be performed. This can take a few minutes.
During this time no measurements are possible. The result of the system check is then
displayed in the overview window.
If the result of one or more of the tests is rated
“Fail,” restart the instrument and repeat the
system check. If the problem persists, please contact the manufacturer or your local
distributor.
6.3
Update
We suggest that you perform a backup before each update (see Section 6.1.12), even if the
update has no influence on the data. In order to backup customer-defined user methods, we
recommend that they are exported to an external storage medium, such as a USB drive.
Download the update file from the MACHEREY-NAGEL homepage and extract the ZIP file
on your PC. Copy the .mns update file to the top level of a USB stick and connect it to the
instrument. Press the
icons and select the required update file from the
list. Wait until a message appears indicating the termination of the update process. (This
process can take up to a few minutes). Restart your device when prompted. Wait until the
unit has restarted and has conducted the necessary self-tests. Please do not turn off the
device during this process.
Press the
icons to see whether the update was successful. The installed
version will now be shown.
6.3.1
Reset
A system reset sets the device back to the default setting. All data, special methods and
custom settings will be deleted. We suggest that you backup all saved data, special
methods, and the user-defined settings beforehand on an SD card (see Section 6.1.12).
After pressing the
icon, a dialogue with a confirmation prompt will open (see Figure 77).
Confirming with
“Yes” will reset the instrument to the default settings. The data stored on the
internal SD card will remain unaffected by the system reset.
Figure 77: System reset
6.3.2
Service
Use of the service menu is protected by a service code. The use of this menu is reserved
exclusively for service technicians for servicing purposes.