IQC
70
NANOCOLOR
®
UV
/
VIS
II and
NANOCOLOR
®
VIS
II V 1.1 / 10.2017
Figure 85: Options of inspection equipment monitoring
6.6.1.1
Lamp check
The lamp check serves to ensure that the installed lamp(s) are working properly. When you
press the
icon on the
NANOCOLOR
®
UV
/
VIS
II, a window will open with a selection of the
lamps to check (see Figure 86).
Please note: This selection is omitted in the user guide of the NANOCOLOR
®
VIS II. The
halogen lamp is checked directly.
Figure 86: Selection of lamps in the lamp check menu
After selecting the lamp and confirming with
, the measuring process will start. The
instrument takes a lamp curve and compares it with the stored reference values. This is
followed by an evaluation of the test. Via the
icon, the result of the test can be printed as
a certificate or stored in the form of a CSV- and PNG-file on an external storage medium.
Summary of Contents for NANOCOLOR UV/VIS II
Page 1: ......