IQC
72
NANOCOLOR
®
UV
/
VIS
II and
NANOCOLOR
®
VIS
II V 1.1 / 10.2017
Figure 88: Running the NANOCHECK program
After all test solutions have been measured, the measured values will be displayed in the
evaluation table to the right of the respective wavelength (see Figure 89). If the value falls
within the confidence interval, it will be displayed in green, otherwise the value will be
displayed in red and the entire measurement
will be given a “Failed” rating. If all values are
within the specified confidence intervals, the test will be given a "Passed" rating.
Figure 89: Results of the screening using the
NANOCONTROL
NANOCHECK
Via the
icon, the result of the test can be printed as a certificate or stored in the form of a
CSV- and PNG-file on an external storage medium.
If the results of the check exceed or fall below the allowed tolerances given in the certificate,
please contact the manufacturer or your local distributor.
6.6.1.3
Wavelength accuracy test
The wavelength accuracy test checks the accuracy of the wavelength at 360 nm, 453 nm
and 536 nm. For measuring purposes, an internal holmium oxide filter is placed into the