58
73VR2102 / 73VR2104 / 73VR2106 / 73VR2108 / 73VR2110 / 73VR2112 USERS MANUAL EM-7395-B Rev.24
Calibration procedure
1. Touch one of the right side fields next to each selection, and enter the true value (expected value) in integer.
Zero
Max. 7 digits including minus sign
Span (+)
Max. 7 digits including minus sign
Span (–)
Max. 7 digits including minus sign
2. Touch OK.
Next to ‘Before/After calibration’ the actual input value measured at the moment of calibration is indicated. The
right side field shows the same value after this particular calibration step. The value is updated in 500 millisecond
cycles.
Touch Cancel to go back to the previous view without applying the calibrated value.
3. Go to the next calibration in the same manner.
4. In order to go to another channel, touch Previous or Next key, or touch Return key to choose one directly on the
view.
Once you have returned to the view to select channels, those already calibrated are indicated with CAL.
Calibrated values are stored in the memory when Record key is touched. Be sure to touch the key and return to
Record view after calibration.
5.7 CONFIRMING CHANGES
New setting becomes valid when the display has gone back to one of the Display views by touching Record key If
there was any change in the storing rate, the data storing form, and/or the number of enabled pens, the message as
shown below appear on the screen.
Touching OK overwrites the previous data file, and touching Cancel return the screen to Display view.
5.8 SETTING UP WITH THE 73VR21BLD
A software configuration created on the 73VR21x Configuration Builder (model: 73VR21BLD) can be downloaded to
the 73VR21x. The configuration set up on the 73VR21x can be uploaded and displayed on the 73VR21BLD.
In order to apply a new setting stored in the CF Card using the 73VR21BLD, restart the 73VR21x. For detailed infor-
mation about the 73VR21BLD, please refer to the users manual for the model.