31
73VR2102 / 73VR2104 / 73VR2106 / 73VR2108 / 73VR2110 / 73VR2112 USERS MANUAL EM-7395-B Rev.24
5.2.2 STORING SETTING
Five storing methods are selectable.
Touching the current selection of the Storing interval shows selectable options at the bottom of the screen. Choose
among the following options:
No storing
No recording
Data is plotted on the chart or displayed on the digital meter or bargraph,
but no data is stored in the CF Card.
Normal
Normal storing mode
Recording is manually initiated and stopped. Data is continuously stored
while the recording is on.
Remote trigger
Remote trigger recording
mode
Data is automatically recorded while the external trigger condition (analog
or discrete input) is true. (See “5.2.3 REMOTE TRIGGER RECORDING”)
Event recording Event recording mode
The 73VR21x detects an external event by trigger signal, and stores pre-
set number of samples (max. 1200 respectively) before and after the mo-
ment of event. (See “5.2.4 EVENT RECORDING”)
Time specified
Store at defined time mode Recording is automatically initiated and stopped at predefined time. (See
“5.2.5 STORE AT A DEFINED TIME MODE”)