54
73VR2102 / 73VR2104 / 73VR2106 / 73VR2108 / 73VR2110 / 73VR2112 USERS MANUAL EM-7395-B Rev.24
n
FVALUE CALCULATION
Reset
In order to reset the totalized value, ‘Trigger input’ and/or ‘Analog input’ can be specified. Touch the ‘None’ key in Page
3/7 for ‘Trigger,’ the one in Page 4/7 for ‘Analog input’ conditions. If you leave the setting to ‘None,’ the data is reset
only when ‘Start’ key is touched.
For details of the trigger input selections, please refer to ‘Peak Hold’ function.
For the analog input resetting, choose one of the following conditions:
None
No resetting by analog input. Reset only when ‘Start’ key is touched.
Value < Limit
Reset when the measured value is lower than the limit value.
Value ≤ Limit
Reset when the measured value is equal to or lower than the limit value.
Limit / Deadband
Specify Limit value used to reset the function result.
Deadband is used to prevent repeating of F value cal-
culation execution and resetting when the measured
value stays unstable close to the limit value.
A positive deadband is used to reset at ‘Limit – Dead-
band,’ while a negative deadband is used to reset at
‘Limit + Deadband.’
By choosing an wide deadband, you can set different
values for set and reset conditions.
Input (X1)
Touching Inp.1 opens the Tag name selector view.
Select an input signal used for the operating function.
Input (X2)
Touching Inp.1 opens the Tag name selector view. Select an input signal used as resetting limit.
Ref. Temp (T0), Z value (Z)
Ref. Temp, Z value
Max. 6 digits including a decimal point and minus sign
Caution !
100 msec. storing interval is not usable with F value calculation.
Be careful to choose a pen of Not Itself. For example, you cannot choose Function Pen 1 or the X2
in an equation for Function Pen 1.
‘e’ can be used to input an exponential value such as ‘1e9.’ Entering ‘e’ in any other way (e.g. ‘1ee’)
will not be recognized as a numeral.
START
RESET
Function value
Measured value
No Start/Reset
Reset !
No Reset
Deadband (positive)