- 40 -
LX400+ UV LED Spot Curing System
Pin 13 INTIN3
Analog input from 1.0VDC to 4.0VDC will set the output
intensity of channel 3 from 15% to 100% in 1%
increments. The following equation can be used to
determine the output intensity based on a voltage input;
y=(85x-40)/3
where;
x is the voltage input from 1.0V to 4.0V.
y is the output intensity in percent. Value round to
nearest percent.
Pin 14 INTIN4
Analog input from 1.0V to 4.0V will set the output
intensity of channel 4 from 15% to 100% in 1%
increments. The following equation can be used to
determine the output intensity based on a voltage input;
y=(85x-40)/3
where;
x is the voltage input from 1.0V to 4.0V.
y is the output intensity in percent. Value round to
nearest percent.
Pin 15 AGND
Return for input analog signals.
Pin 21 EMPTY
Not In Use.