- 39 -
LX400+ UV LED Spot Curing System
Pin 7 SYNCOUT3 (+) and Pin 19 SYNCOUT3(-)
Signal output to indicate LED output of channel 3 status.
When the LED head is on, the signal is active. When the
LED head is off, the signal is inactive.
Pin 8 SYNCOUT4 (+) and Pin 20 SYNCOUT4(-)
Signal output to indicate LED output of channel 4 status.
When the LED head is on, the signal is active. When the
LED head is off, the signal is inactive.
Pin 9 EMPTY
Not In Use.
Pin 10 STRT/STP1
Shorting signal to GND will trigger channel 1.
STRT/STP1 = 0V. Channel 1 triggers.
STRT/STP1 = floating. No change.
Pin 11 STRT/STP2
Shorting signal to GND will trigger channel 2.
STRT/STP2 = 0V. Channel 2 triggers.
STRT/STP2 = floating. No change.
Pin 12 FPLOCK
Shorting signal to GND will lock front panel.
FPLOCK = 0V. Locks front panel.
FPLOCK = floating. No change.