![Lightmed LightLink-CXL Service Manual Download Page 86](http://html1.mh-extra.com/html/lightmed/lightlink-cxl/lightlink-cxl_service-manual_1912191086.webp)
LightLink-CXL – Service Manual
Rev. No. : 01
Page
85
of
98
Section 10
Software Upgrading Procedures
10.1
Introduction
The purpose of this procedure is to upgrade or rewrite the software for cross linking
system. Performing this procedure will erase the original program and install the new
version into the CPUs for the whole system.
10.2
Preparations
Equipment Requirement
-MSP-FET430UIF kit (including MSP430 CD-ROM, MSP-FET430UIF USB-to-JTAG
Emulation Interface, USB cable type A/B, 14-conductor cable)
-PC or Laptop
System Requirement
Window XP or above
10.3
Driver Software Downloading Procedures
- Access the website
http://www.elprotronic.com/download.html
or “Flash Programmers for TI's MSP430 MCU uses TI's FET Adapter” (refer to
below figure)
Note: This is freeware software therefore the version could be different from time to
time
Summary of Contents for LightLink-CXL
Page 1: ...Service Manual LightLink CXL Corneal Cross Linking System ...
Page 90: ...LightLink CXL Service Manual Rev No 01 Page 89 of 98 ...
Page 91: ...LightLink CXL Service Manual Rev No 01 Page 90 of 98 ...
Page 93: ...LightLink CXL Service Manual Rev No 01 Page 92 of 98 ...
Page 97: ...LightLink CXL Service Manual Rev No 01 Page 96 of 98 Finish this process we will get PASS ...
Page 99: ...LightLink CXL Service Manual Rev No 01 Page 98 of 98 ...