![Lightmed LightLink-CXL Service Manual Download Page 50](http://html1.mh-extra.com/html/lightmed/lightlink-cxl/lightlink-cxl_service-manual_1912191050.webp)
LightLink-CXL – Service Manual
Rev. No. : 01
Page
49
of
98
B. Aiming Module Replacement
Removal
1.
Remove the laser head cover off (both the front and rear cover)
2.
Disconnect the aiming connector (Depending on the which one)
3.
Loosen the screw by using 3.0 mm Allen key
4.
Gently and slowly remove the laser diode and holder
Note: Do remember to tag all the connectors
Figure 6.3 Aiming set screw
Figure 6.4 Aiming connectors
Two aiming connectors
Set screw
Summary of Contents for LightLink-CXL
Page 1: ...Service Manual LightLink CXL Corneal Cross Linking System ...
Page 90: ...LightLink CXL Service Manual Rev No 01 Page 89 of 98 ...
Page 91: ...LightLink CXL Service Manual Rev No 01 Page 90 of 98 ...
Page 93: ...LightLink CXL Service Manual Rev No 01 Page 92 of 98 ...
Page 97: ...LightLink CXL Service Manual Rev No 01 Page 96 of 98 Finish this process we will get PASS ...