LightLink-CXL – Service Manual
Rev. No. : 01
Page
51
of
98
C.
Detector PCB Replacement
Removal
1.
Remove the laser head cover off (only the front cover)
2.
Disconnect and tag the cablings or connectors
3.
Remove the three screws (refer to CPU replacement procedures)
4.
Gently and slowly remove the PCB down
Note: Do remember to tag all the connectors
5.
Remove the cable connection and the two screws by using 1.5mm Allen key
6.
Gently and slowly remove the PCB down
Figure 6.5 Detector PCB replacement
Installation
1.
Reseat the PCB back and screw back the two screws
2.
Reconnect back the connector
3.
Reseat the CPU PCB back to the screw pole and screws back the three screws
4.
Reconnect all the connectors and verify them carefully prior to power on
Note: Do remember to reconnect all connectors before turning on the power
5.
Ensure there is no loosen connections and close up the cover top
6.
Power on and validate the system performance refer to section 4.6 on page 31
7. Recalibrate the Power and Detector parameter (refer to recalibration
procedures)
Two 1.5mm screws
Detector Connector
Summary of Contents for LightLink-CXL
Page 1: ...Service Manual LightLink CXL Corneal Cross Linking System ...
Page 90: ...LightLink CXL Service Manual Rev No 01 Page 89 of 98 ...
Page 91: ...LightLink CXL Service Manual Rev No 01 Page 90 of 98 ...
Page 93: ...LightLink CXL Service Manual Rev No 01 Page 92 of 98 ...
Page 97: ...LightLink CXL Service Manual Rev No 01 Page 96 of 98 Finish this process we will get PASS ...
Page 99: ...LightLink CXL Service Manual Rev No 01 Page 98 of 98 ...