LightLink-CXL – Service Manual
Rev. No. : 01
Page
62
of
98
B. Power Calibration (Power Parameter)
This procedure will calibrate the power performance which allows the system to have
the corrected power throughput during the treatment procedure. Improper power will
affect the treatment quality.
- Enter to calibration mode
- Double clicks the hidden icon at the left top corner screen (refer to Figure 7.1)
- Press the "Power Parameter" icon to initial the calibration process
- Set the power meter wavelength at 365 nm and set the spot size to 11mm in main
calibration menu (refer to motor parameter) and power to 30 mW
- Initial ‘Ready’ button for spot size reinitialization or reconfiguration process
before entering the Power Parameter menu
Figure 7.9 Wavelength Setup
Figure 7.10 Power Parameter Menu
Wavelength at
Press 30 mW
Set
'Power' at
30mW