LightLink-CXL – Service Manual
Rev. No. : 01
Page
77
of
98
-
Identify which aiming spot position correspondence for VR
- Use VR pot adjuster to adjust the value to 450
W
for both two aiming (Refer
to below figure)
Figure 7.23 VR Location For Aiming Energy
- Repeat the same process until the value is reached
Aiming Power Calibration Record Sheet
System SN:
Power Meter & Detector:
Performed by:
Dated:
Item
Aiming A
Intensity Scale
Output A
Power
Power Meter
Readout
Aiming B
Intensity Scale
Output B
Power
Power Meter
Readout
1
10
425~475μW
10
425~475μW
2
1
25~75μW
1
25~75μW
Two VR control aiming energy