LightLink-CXL – Service Manual
Rev. No. : 01
Page
69
of
98
D. Spot Size Power Calibration (Power Spot Parameter)
This procedure is intended to calibrate the individual spot sizes power for the system
spot power consistency and accuracy during the treatment procedure.
- Enter to calibration mode
- Press the "Power Spot Parameter" icon to initial the calibration process
- Set the power meter wavelength at 365nm and set the spot size to 11mm in main
calibration menu (refer to motor parameter) and power to 30 mW
- Start UV light run icon and measure if the value falls within the calibration
specification guideline(Refer to below figure)
Figure 7.16 Power Spot Parameter