LightLink-CXL – Service Manual
Rev. No. : 01
Page
80
of
98
Section 8
MAINTENANCE
The LightLink-CXL Corneal Cross-Linking System has been designed to require minimal
maintenance. There are several simple routine procedures that are to be carried out by
the operator, but aside of these there are no operator maintenance requirements.
The manufacturer however does recommend that the LightLink-CXL be serviced by an
authorized service agent every 12 months. During this Preventative Maintenance (PM)
service visit the UV light System will be Calibrated and Aligned and the general operating
function confirmed.
It is a requirement that on an annual basis (every 12 months) the LightLink-CXL has the
Power meter calibrated to a known calibrated meter and have its earth leakage current
and earth resistance measured according to EN60601-1. These procedures can only be
performed by an authorized service agent of the manufacturer.
If at any time you have concerns about any aspect of the Operation / Calibration or
Alignment of the LightLink-CXL, you are urged to contact the authorized representative
or the manufacturer in order to decide on a suitable course of action.
Refer to Section 9 of this Manual for particular National requirements of EU Countries.
8.1
Operator / User Maintenance
It is recommended that the
LightLink-CXL Corneal Cross-Linking System
equipment is
annually checked by LightMed Authorized Representative. Please contact the Authorized
Service to check the technical and commercial conditions.
LightMed Corp. cannot be held liable for the equipment if it is opened and/or
modified by unauthorized persons.
If the equipment is used in a manner other than that specified by the Engineering,
the warrantee will automatically be cancelled. Read carefully the operation
conditions. If in doubt, see our Technical Assistance Service.
8.2
Cleaning
The cleaning should be done with a lint-free cloth moistened with a non-abrasive
cleaning solution. For example, water and a mild detergent are more appropriate. Do not
use organic solvents as gasoline, alcohol, kerosene, etc. for cleaning the outside of the
LightLink-CXL.
When cleaning the inside of the dome of the stimulator it is advisable to use a special
soft cleaning cloth to reduce the possibility of scratching the LCD display.
Summary of Contents for LightLink-CXL
Page 1: ...Service Manual LightLink CXL Corneal Cross Linking System ...
Page 90: ...LightLink CXL Service Manual Rev No 01 Page 89 of 98 ...
Page 91: ...LightLink CXL Service Manual Rev No 01 Page 90 of 98 ...
Page 93: ...LightLink CXL Service Manual Rev No 01 Page 92 of 98 ...
Page 97: ...LightLink CXL Service Manual Rev No 01 Page 96 of 98 Finish this process we will get PASS ...
Page 99: ...LightLink CXL Service Manual Rev No 01 Page 98 of 98 ...