7.2
Alignment Procedures
Purpose
The purpose of this procedure is to align or restore the laser system after any module
replacement or during any routine checkout process. Prior to the commencing the
procedure, only the qualified personnel is permitted to conduct this process.
There are two alignments required when replacing:
1. Aiming Alignment
- Intensity
- Overlapping
2. CCD Camera Alignment
Equipments / Tools required:
Interlock Connector
Calibrated Laser Power Meter
Safety Goggle (365nm wavelength)
VR pot adjuster or small flat screw driver
Allen key 2.0 mm