Spot Size Power Record Sheet
System SN:
Power Meter & Detector:
Performed by:
Dated:
Item
Power
Interval
Spot Size
Standard Output
Power
Power Meter
Readout
Spot Size
Factor
Pass(V) Fail(X)
1
45mW
11mm
42.8mW±20%
2
45mW
10mm
35.3 mW±20%
3
45mW
9mm
28.6mW±20%
4
45mW
8mm
22.6mW±20%
5
45mW
7mm
17.3mW±20%
6
45mW
6mm
12.7mW±20%
7
45mW
5mm
8.8mW±20%
8
45mW
4mm
5.7mW±20%