LightLink-CXL – Service Manual
Rev. No. : 01
Page
16
of
98
Section 4
SYSTEM SETUP
4.1
Introduction and Requirements
It is strongly recommended that the manufacturer or its authorized agent install the
LightLink-CXL Corneal Cross-Linking System at the operator site to ensure that the
system is operating correctly, aligned and calibrated according to specification. After
this initial installation it is the operator’s responsibility to ensure the UV light system is
operating correctly whenever the UV light is moved or relocated.
The following procedures should be followed in order to successfully install the UV light
system. The checklist and report form should be completed and a copy sent to the
manufacturer.
In the event that the report is not sent to the manufacturer then the
manufacturer reserves the right to decline any warranty claims that may be
forthcoming.
The installer should also retain a copy and the customer may keep a copy too.
All precautions care must be taken when installing the LightLink-CXL Corneal Cross-
Linking System to ensure that you or others are not exposed to any hazardous UV light
radiation. Always wear UV light safety glasses suitable for this particular wavelength
to protect your eyes.
The installation requirements are:
1.
100-230 Volts, 50 or 60Hz AC mains power supply with an earth connection. This is a
single-phase outlet capable of delivering up to 400 Watts.
2.
A mains power cable is supplied but the connector may not suit the outlet available
so it is advisable to have a spare locally compatible cable available. The Cable
assembly must be CE approved for EU Countries.
3.
A suitable room to place the UV light System in that provides for a safe working
environment is required. As with other ophthalmic equipment a dimly lit room is
preferred.
4.
The UV light System has the facility to connect a remote door interlock to the UV
light treatment room. If this option is required then the customer must organize this
with an electrician and the manufacturer or authorized agent can provide
instructions on how to connect to the UV light System. The UV light System is
provided with a Bypass plug in the event that this option is not installed. This plug is
referred to as the “Interlock Key” and must be inserted into the control box in order
for the system to operate. Do not remove this Bypass connector unless you intend
to install the remote door interlock switches at the site. Removing the connector
will prevent the UV light from operating.
5.
If the user needs to reallocate the system to a new location it is recommended to
unlock all the movement stoppers on the base wheel station and carefully transport
Summary of Contents for LightLink-CXL
Page 1: ...Service Manual LightLink CXL Corneal Cross Linking System ...
Page 90: ...LightLink CXL Service Manual Rev No 01 Page 89 of 98 ...
Page 91: ...LightLink CXL Service Manual Rev No 01 Page 90 of 98 ...
Page 93: ...LightLink CXL Service Manual Rev No 01 Page 92 of 98 ...
Page 97: ...LightLink CXL Service Manual Rev No 01 Page 96 of 98 Finish this process we will get PASS ...
Page 99: ...LightLink CXL Service Manual Rev No 01 Page 98 of 98 ...