Chapter 2: Quality Control Test
19
When to Perform a Quality Control Test
• As required by your institution’s quality control
policy or local regulatory requirements.
• When you open a new bottle of test strips.
• If a patient test has been repeated and the blood
glucose results are still lower or higher than
expected.
• After cleaning the meter.
• If there are other indications that the system is not
working properly.
• If you drop the meter.
QC enforcement
❍
Warning only
❍
Meter lockout
To ensure that QC testing is performed on a meter
as required, a warning mode or lockout mode was
selected. Both modes display a message informing
you that QC testing is due. The warning mode
allows you to continue with a patient test; the
lockout mode prevents you from performing a
patient test or special tests until QC is successfully
performed.
❑
Require operator QC every ___ days
A second lockout mode prevents each operator who
has not performed QC testing within the
designated number of days from using the meter to
test patient blood.
Stealth.book Page 19 Monday, October 28, 2002 11:20 PM
Summary of Contents for SureStepFlexx
Page 1: ......
Page 2: ......
Page 8: ...vi...
Page 13: ...1 Introduction...
Page 28: ...16 SureStepFlexx Meter Operator s Guide...
Page 29: ...2 Quality Control Test...
Page 41: ...3 Patient Test...
Page 51: ...4 Data Transfer and Management...
Page 61: ...5 Special Tests...
Page 74: ...62 SureStepFlexx Meter Operator s Guide...
Page 75: ...6 Care and Maintenance...
Page 89: ...7 Troubleshooting...
Page 108: ...96 SureStepFlexx Meter Operator s Guide...
Page 116: ...104 SureStepFlexx Meter Operator s Guide...
Page 117: ......