Appendix A: Meter Configuration Options
93
Operator and IDs
Configuration Option
Meaning
❑
Operator ID required
for all tests
❑
Maintain ID for
___ minutes after
power off
❑
Require
operator QC
every ___ days
If selected all operators are
required to enter their IDs for
each test.
Maintain ID remembers the last
ID for 1 to 60 minutes from the
time the meter is turned off until
it is turned on again.
Require operator QC ensures each
operator who uses the meter
performs the required QC tests.
ID verification
❑
Check for non-validated
operators
❑
Check for non-
certified operators
❍
Warning only
❍
Lockout
Warn or lock out non-validated
operators prevents operators who
are not in the system from using
the meter.
Warn or lock out non-certified
operators prevents operators who
are not certified or whose
certification has expired from
using the meter.
❑
Patient ID required
If selected, patient IDs are
required.
ID format string
(operator and patient)
Format strings were defined for
all operator and patient IDs that
will be entered at the meter. The
string defines the number and
type of characters allowed.
Operator re-certification
setting
Period, ___ months
___ High control tests
___ Low control tests
___ Normal control tests
___ Patient tests
The period is the frequency at
which each operator must be
re-certified. Test settings define
the number of high, low,
normal, and patient tests an
operator must run within the
certification period to be eligible
for re-certification.
✔
✔
Stealth.book Page 93 Monday, October 28, 2002 11:20 PM
Summary of Contents for SureStepFlexx
Page 1: ......
Page 2: ......
Page 8: ...vi...
Page 13: ...1 Introduction...
Page 28: ...16 SureStepFlexx Meter Operator s Guide...
Page 29: ...2 Quality Control Test...
Page 41: ...3 Patient Test...
Page 51: ...4 Data Transfer and Management...
Page 61: ...5 Special Tests...
Page 74: ...62 SureStepFlexx Meter Operator s Guide...
Page 75: ...6 Care and Maintenance...
Page 89: ...7 Troubleshooting...
Page 108: ...96 SureStepFlexx Meter Operator s Guide...
Page 116: ...104 SureStepFlexx Meter Operator s Guide...
Page 117: ......