Chapter 6: Care and Maintenance
65
Cleaning the Test Strip Holder and Lens
To clean the test strip holder (cover and base), lens
area, and contact points, use a 10% bleach solution
followed by water. Dry thoroughly.
▲
CAUTION:
Do not use alcohol, glass cleaners, or any
cleansers containing abrasives, phenol, or ammonia to
clean the test strip holder or lens area because it will
damage the meter parts.
1
Press down on the left side of the test strip
holder. This releases the holder, allowing you
to slide it from the meter.
2
Wipe the test strip holder cover and base with
a cotton swab or soft cloth dampened with a
10% bleach solution. Be sure to thoroughly
wipe the gray area on the inside cover. Clean
both sides of the base. Follow with a cotton