52
SureStepFlexx Meter Operator’s Guide
5
Select the linearity solution
lot number from the list
displayed.
The last lot number selected
appears at the top of the list.
Use the arrow(s) to scroll
through the list, if necessary.
If the meter is equipped with
the optional barcode scanner, you may press
Entr Lot#
and scan the barcode on the linearity solution vial.
6
Select the test strip lot
number (and code) from
the list displayed.
The last test strip lot number
selected appears at the top of
the list. Use the arrow(s) to
scroll through the list, if
necessary.
❑
Require solution lot number
The Select Lin Lot screen appears only if you are
required to enter the lot number.
❑
New reagent entry at workstation only
The meter may be set to allow you to enter new
reagent information. If the linearity solution lot
number does not appear in the list, press the
Entr Lot#
button and enter the linearity solution
lot number and expiration date.
S e l e c t L i n L o t
M G 2 0 7 5
M L 3 3 4 7
M G 9 0 5 6
L S 4 7 8 3
L o t #
Entr
Lot#
Stealth.book Page 52 Monday, October 28, 2002 11:20 PM
Summary of Contents for SureStepFlexx
Page 1: ......
Page 2: ......
Page 8: ...vi...
Page 13: ...1 Introduction...
Page 28: ...16 SureStepFlexx Meter Operator s Guide...
Page 29: ...2 Quality Control Test...
Page 41: ...3 Patient Test...
Page 51: ...4 Data Transfer and Management...
Page 61: ...5 Special Tests...
Page 74: ...62 SureStepFlexx Meter Operator s Guide...
Page 75: ...6 Care and Maintenance...
Page 89: ...7 Troubleshooting...
Page 108: ...96 SureStepFlexx Meter Operator s Guide...
Page 116: ...104 SureStepFlexx Meter Operator s Guide...
Page 117: ......