92
SureStepFlexx Meter Operator’s Guide
Meter
Configuration Option
Meaning
❑
Meter lockout if no
upload in ____ days
❑
Warn prior to
lockout
If selected, the meter cannot be
used for testing if a data transfer
session has not been performed
within the specified number of
days (maximum of 62 days).
A warning may appear 2 days
prior to the upload date.
❑
Enable barcode
scanner
If selected, the meter barcode
scanner is enabled. If not
selected, you must enter all IDs
and lot numbers manually or
select from the list displayed.
Location to which
meter is assigned
Each meter is assigned to a
particular location. Do not use
the meter outside the designated
location.
❑
New reagent entry at
workstation only
If selected, you must choose an
existing test strip or control lot
number from the list displayed. If
not selected, you may enter new
test strip and control solution lot
numbers that have not yet been
entered in the system.
❑
Require solution lot
number
If selected, operators must enter
valid control, linearity, and
unknown solution lot numbers
when performing those tests.
Modem dialing
information
The receiving modem phone
number can be preprogrammed.
If preprogrammed, you do not
need to enter the phone number
at the meter when transferring
data.
✔
✔
Stealth.book Page 92 Monday, October 28, 2002 11:20 PM
Summary of Contents for SureStepFlexx
Page 1: ......
Page 2: ......
Page 8: ...vi...
Page 13: ...1 Introduction...
Page 28: ...16 SureStepFlexx Meter Operator s Guide...
Page 29: ...2 Quality Control Test...
Page 41: ...3 Patient Test...
Page 51: ...4 Data Transfer and Management...
Page 61: ...5 Special Tests...
Page 74: ...62 SureStepFlexx Meter Operator s Guide...
Page 75: ...6 Care and Maintenance...
Page 89: ...7 Troubleshooting...
Page 108: ...96 SureStepFlexx Meter Operator s Guide...
Page 116: ...104 SureStepFlexx Meter Operator s Guide...
Page 117: ......