LI-600 Porometer/Fluorometer Instruction Manual
9-2
Figure 9-1
. Air flow measured before and after interaction
with a leaf.
Relative humidity (RH) sensors on the cuvette measure
the air stream before and after it interacts with a leaf.
Flow rates are measured before and after passing
through the cuvette; leaf temperature is measured in the
cuvette.
Derivation of equations
The mass balance of water vapor in an open system at
steady-state is given by
9-1