Section 8. Maintenance
8-5
Figure 8-4
. Shell removal allows you to perform internal
maintenance.
Removing and inserting a battery
Removing and inserting the battery is only required
when the battery no longer functions and needs replace-
ment.
1
Prepare the LI-600.
Power off the instrument and ground yourself to
avoid potential electrostatic discharge.
2
Remove the shells.
Remove the shells on the instrument head. Set the
shells safely aside (see
3
Unplug the battery connector.
On the right side of the instrument head, locate the
cable clip. Push down on the clip and disconnect the
battery connector.
4
Remove the coverings on the instrument handle.
Remove the six screws nearest the USB port on the
handle using a Phillips screwdriver; do not remove
the two hex screws at the end of the handle. Gently
spread apart the left and right handle coverings but
do not remove completely.
5
Detach the endcap.
The dust cap may dislodge while removing the end-
cap. Set both safely aside.
6
Remove the old battery.
To the right of the connector slot, ease the wires out
from behind the tab. Gently remove the old battery