LI-600 Porometer/Fluorometer Instruction Manual
7-2
Caution:
Do not apply pressure to the aperture
clamp to achieve a seal on the leaf. This can warp the
aperture clamp, interfere with the instrument’s
optics, and produce measurement errors.
Instrument will not automatically log a
measurement
If the instrument will not automatically log a meas-
urement, check that Auto Mode is enabled for the given
configuration (see
Ensure that the instrument is properly sealing on a leaf.
Ensure that stability criteria are adequate for your meas-
urement. See
Stability criteria in auto mode
The display layout is blank
A blank layout will appear on the display if no variables
have been selected for the configuration. A meas-
urement can still be taken if the display layout is blank.
Selected the desired variables for the layout in con-
figuration management in the software (see
A blank layout will appear on the display if a con-
figuration was removed from the instrument before exit-
ing a corresponding menu. Upload the configuration
using the computer software then select it on the instru-
ment.
Software warning when connecting the LI-600
If the computer software displays a warning message
when connecting the LI-600, the firmware version or
computer software version may require updates.